Chapter 4 Command Reference
:CALCulate[1] Subsystem
4-10 Programming Guide
:CALCulate[1]:MATH[:EXPRession][:DEFine]?
Queries equation used for math operation.
A sequence of ASCII-encoded bytes:
(SENS * SCALE + OFFSET)
terminated with a new line and EOI.
Query only.
This query should be the last query in a terminated program message;
otherwise, error -440 is generated.
:CALCulate[1]:MATH[:EXPRession]:NAME SCALE_OFFSET
or
:CALCulate[1]:MATH[:EXPRession]:SELect SCALE_OFFSET
Sets or queries the name of the expression selected for math processing .
A sequence of ASCII-encoded bytes: SCALE_OFFSET
*RST: SCALE_OFFSET
:CALCulate[1]:MATH:STATe <Boolean>
Sets or queries the math enable.
This enable specifies whether or not measurement (SENSe) data will be scaled and
offset.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: OFF
Updating the math enable causes the limit counts (:CALC2:LIM:FCO,
:CALC2:LIM:PCO) to be cleared.