HP 8568B Uživatelský manuál Strana 142

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 313
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 141
1. Low-Voltage Power Supply Adjustments
supply, therefore, if the +5V supply is out of tolerance, a circuit
malfunction is indicated.
17. The
-5V
indicator
A24DS6
(yellow LED) should be lit.
18. Connect the DVM to
A24TP7.
The DVM indication should be
-5.200
ho.050
V
de.
The
-5V
supply is referenced to the
+2OV
supply, therefore, if the
-5V
supply is out of tolerance, a circuit
malfunction is indicated.
19. The
-15V
indicator
A24DS3
(yellow LED) should be lit.
20. Connect the DVM to
A24TP4.
The DVM indication should be
-15.000
f0.050
V dc. The
-15V
supply is referenced to the +
20V
supply, therefore, if the
-15V
supply is out of tolerance, a circuit
malfunction is indicated.
3-28 Adjustments
Zobrazit stránku 141
1 2 ... 137 138 139 140 141 142 143 144 145 146 147 ... 312 313

Komentáře k této Příručce

Žádné komentáře