ARM
The ARM subsystem selects the number of scanning cycles (1 through
32767) for each
INITiate command.
Subsystem Syntax ARM
:COUNt <
number> MIN | MAX
:COUNt? [MIN | MAX]
:COUNt ARM:COUNt <number> MIN | MAX allows scanning cycles to occur a
multiple of times (1 to 32767) with one
INITiate command when
INITiate:CONTinuous OFF |0 is set. MIN sets 1 cycle and MAX sets 32767
cycles.
Parameters
Parameter Name Parameter Type Range of Values
<number>
numeric 1 |32767 |MIN |MAX
Comments Number of Scans: Use only values between 1 to 32767 for the
number of scanning cycles.
Related Commands: ABORt, INITiate:IMMediate
*RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles
This example sets an RF multiplexer for 10 scans of channels 00 through
03. When the scan sequence completes, channel 03 is closed.
ARM:COUN 10 ! Set 10 scans per INIT command
SCAN (@100:103) ! Scan channels 00-03
INIT ! Start scan, close channel 00
Chapter 5 HP E1366A/E1367A RF Multiplexers Command Reference 41