Preparing For Use
Power Requirements
Checking
the
Fuse
Figure 1-3. Checking the Line Fuse
The recommended fuse is size 5 by 20 mm, rated
F5A,
250 V (IEC approved).
This fuse may be used with input line voltages of 115 V or 230 V. Its HP part
number is 2110-0709.
With an input line voltage of 115 V an alternate fuse can be used. In areas
where the recommended fuse is not available, a size 5 by 20 mm, rated fast
blow, 5 A, 125 V
(UL/CSA
approved) fuse may be substituted. Its HP part
number is
2110-0756.
The line fuse is housed in a small container beside the rear-panel power
connector (see Figure l-3). The container provides space for storing a spare
fuse, as shown in the figure.