HP Data Explorer 4 Series Uživatelský manuál Strana 331

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 447
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 330
Voyager Data Examples
Data Explorer
Software Users Guide 7-15
7
Before calibrating To optimize mass accuracy, do the following before calibrating:
1. Display the spectrum of interest.
Baseline correcting 2. From the Process menu, select Baseline Correction.
The spectrum is baseline corrected. For more
information, see Section 5.8.2, Using Baseline
Correction.
Deisotoping 3. From the Peaks menu, select Peak Deisotoping.
The Deisotoping dialog box (Figure 7-14) is displayed.
Figure 7-14 Deisotoping Dialog Box
4. For this example spectrum, specify H for Adduct and
C6H5NO for Generic Formula.
5. Click OK.
Figure 7-15 shows the baseline-corrected, deisotoped
spectrum before calibration. For more information on
deisotoping, see Section 3.4, Deisotoping a Spectrum.
Figure 7-15 Baseline-Corrected, Deisotoped Spectrum
Before Calibration
Zobrazit stránku 330
1 2 ... 326 327 328 329 330 331 332 333 334 335 336 ... 446 447

Komentáře k této Příručce

Žádné komentáře